Product setup

ISO 80369-20 Annex D Subatmospheric Air Leakage Tester

KingPo supplied documentation identifies this tester for ISO 80369-20 Annex D subatmospheric-pressure air leakage workflows.

KP6920D / ISO80369-20 Annex DAvailable for quotationConfirm calibration scope
Documented model scopeISO80369-20 Annex D
Catalogue modelKP6920D
Applicable AnnexAnnex D
Use whenThe requested ISO 80369-20 route is subatmospheric-pressure air leakage
Annex D workflow

Reduced-pressure air leakage testing with a defined setup chain

This tester is documented for ISO 80369-20 Annex D subatmospheric-pressure air leakage. It combines the vacuum channel, sample-volume input, pressure display, leakage calculation and report controls needed to prepare a repeatable laboratory workflow.

Choose the correct method

Use this equipment when the requested route is Annex D subatmospheric air leakage. Positive-pressure air or liquid leakage should be specified through the separate Annex B/C tester.

Operating sequence
  1. Place the tester on a stable bench, connect power and select the Pressure function.
  2. Install the sample with the required mating reference and fixture support.
  3. Enter test time, reduced-pressure target and sample volume, then start the test.
  4. Adjust the air-pressure valve to the set value and review the curve and result record.

Documented controls and result fields

  • 0 to -90 kPa documented pressure range
  • 0-9999 s settable test time and one-station configuration
  • Starting pressure, pressure change, test period and sample volume
  • Leakage-rate display, curve review and print function

Confirm before quotation

  • Annex D method and connector sample identity
  • Mating reference connector and sample holding fixture
  • Required reduced-pressure condition and test duration
  • Result fields, printout, manual and certificate scope

The operating workflow and core values shown here are taken from KingPo supplied material. The catalogue configuration and the operating manual are identified separately where their level of detail differs.

Public method reference

Technical reference: Annex D subatmospheric test chain

Figure D.1 identifies the vacuum source, measurement channel, test volume, specimen connection and reference connector roles for the Annex D method. Use it to plan the complete test setup; it is not a product photograph or a controlled manufacturing drawing.

Click the drawing to inspect it at a larger size. Request the selected KingPo equipment setup separately.

Key information

Power supply220V 50Hz
Test station1 station
Pressure range0 to -90 kPa
Pressure gauge range0 to -90 kPa
Pressure gauge accuracy+/- 0.25% FS
Accuracy+/- 0.25%
Test time0-9999 s can be set
Control modePLC + 7 inch touch screen
Output and input pipe sizeØ4 mm
Annex selection noteUse Annex D for reduced-pressure air leakage; use the Annex B/C tester for positive pressure or liquid routes
Dimensions360 x 380 x 210 mm (W x D x H)
Weight11.5 kg
Source materialKingPo supplied operating documentation

What to confirm before ordering

Standard / FigureConfirm applicable standard and product family
Product typetest equipment
Connector familyleakage tester family
Information for quotationMating connector, method, quantity, fixture and certificate needs
CalibrationDiscuss certificate scope and traceability needs for the requested product scope
Base configurationFixture set configurationCalibration package configuration

Share the accessory, fixture, certificate and shipping needs so KingPo can identify the right product scope.

Technical documents

Request documents for the exact product configuration

Datasheets, manuals, controlled drawings, report examples, calibration records and certificate wording should match the selected Figure, Annex, fixture, instrument channel and order configuration.

Product imageAvailable on this page
Technical drawingControlled drawing requested when applicable
DatasheetRequest for selected product scope
Request typesDatasheet request, Configuration checklist request
CalibrationCertificate scope discussed for the selected product scope
CertificateCertificate wording confirmed with quotation scope
Test method

How this product fits into the method

Use the test method to decide whether the order needs only this item, a matched counterpart, a fixture, test equipment or a complete set.

Test methodProduct roleTypical additional equipment
Subatmospheric leakageMating reference at the reduced-pressure connection pointVacuum source, leakage measurement channel and connector fixture

RFQ context

Quote requests should include product, Figure, standard part, connector side, mating counterpart, test method, fixture need, quantity, calibration requirement and certificate expectation.

Request this product

FAQ

When should I choose Annex D instead of the Annex B/C leakage tester?

Choose the Annex D tester for subatmospheric-pressure air leakage. Choose the Annex B/C tester for pressure or positive-pressure liquid leakage.

What RFQ details matter for Annex D?

Send the connector sample, mating or reference connector, fixture support, requested pressure route, report expectation and whether the controlled operation manual or certificate wording is needed.