chinatic.org / ISO 80369 Series
ISO 80369-2:2024 · Clause-by-clause reading

ISO 80369-2:2024: Performance Requirements (Clauses 7.1 to 7.7)

Part 5 of 7  ·  Clauses covered: Clauses 7.1 to 7.7

Clause 7 contains seven performance requirements. Each is built from the same four elements: an applied condition, an acceptance criterion, a set of reference connectors identified by figure number in Annex C, and a test method annex in ISO 80369-20 (YY/T 0916.20—2019 in the Chinese draft). Most clauses permit a higher pressure, force or torque, or a longer dwell time, to be applied. None permits a lower one.

ClauseRequirementApplied conditionAcceptanceReference connectorsISO 80369-20 annex
7.1Leakage by pressure decayAir, 30 s to 35 s. R1: 12.5 kPa to 15.0 kPa. R2: 600 kPa to 640 kPa.R1 ≤ 0.00025 Pa·m³/s
R2 ≤ 0.005 Pa·m³/s
Figures C.1, C.3, C.8, C.10B
7.2Sub-atmospheric pressure air leakageR1: 3.0 kPa to 5.0 kPa negative pressure, 25 s to 35 s. R2: 35.0 kPa to 45.0 kPa negative pressure, 20 s to 30 s.R1 ≤ 0.00005 Pa·m³/s
R2 ≤ 0.005 Pa·m³/s
Figures C.1, C.3, C.8, C.10D
7.3Stress crackingStress applied in accordance with Annex E of ISO 80369-20Requirements of 7.1 and 7.2 still met afterwardsFigures C.1, C.3, C.8, C.10E
7.4Resistance to separation from axial load32 N to 35 N, 10 s to 15 sNo separation from the reference connectorFigures C.2, C.4, C.9, C.11F
7.5Resistance to separation from unscrewing0.018 N·m to 0.020 N·m, 10 s to 15 sNo separation from the reference connectorFigures C.1, C.3, C.8, C.10G
7.6Resistance to overriding5 s to 10 s. R1: 0.15 N·m to 0.17 N·m. R2: 0.22 N·m to 0.25 N·m.No overriding of the thread or lugs of the reference connectorFigures C.2, C.4, C.9, C.11H
7.7Disconnection by unscrewingUnscrewing torque not greater than 0.35 N·mSeparation from the reference connector shall be possibleFigures C.1, C.3, C.8, C.10I

7.1 and 7.2. Leakage in both directions

Both connector families are evaluated for fluid leakage by pressure decay, with air as the test medium and a holding time of 30 s to 35 s. The R1 test pressure is 12.5 kPa to 15.0 kPa with a leakage rate not exceeding 0.00025 Pa·m³/s; the R2 test pressure is 600 kPa to 640 kPa with a limit of 0.005 Pa·m³/s. Annex A states that these pressures represent the worst case arising under single-fault conditions for an R1 connector in a breathing system and an R2 connector in a medical gas pipeline system. This is why the test pressures sit at the upper end of the application ranges given in Clause 1.

Sub-atmospheric leakage is a separate requirement, and the asymmetry between R1 and R2 is larger. R1 is subjected to a negative pressure of 3.0 kPa to 5.0 kPa for 25 s to 35 s with a limit of 0.00005 Pa·m³/s, while R2 is subjected to 35.0 kPa to 45.0 kPa for 20 s to 30 s with a limit of 0.005 Pa·m³/s. The method is Annex D of ISO 80369-20, and a higher negative pressure may be applied. Table D.1 in the informative Annex D shows why negative-pressure performance matters: gas sampling lines and fluid sampling lines are listed with pressure ranges down to −400 hPa and −800 hPa respectively.

7.3. Stress cracking as a combined judgement

Clause 7.3 sets no numerical limit of its own. The connector is first subjected to the stress specified in Annex E of ISO 80369-20, using the same reference connectors as the leakage tests, and then has to satisfy 7.1 and 7.2. What is being examined is the integrity of the sealing surfaces after assembly and environmental stress, rather than a separate failure mode. This is also the reason the material requirements of Clause 5 and the reference connector requirements of Annex C are both expressed as modulus thresholds: the specimen must not creep under load, and the datum must not wear while the load is applied.

7.4 to 7.7. Retention, overriding and release

The four mechanical clauses have to be read together because their acceptance criteria point in different directions. Clause 7.4 requires the connection to withstand an axial separation force of 32 N to 35 N for 10 s to 15 s. Clause 7.5 requires it to withstand an unscrewing torque of 0.018 N·m to 0.020 N·m for the same dwell time. Clause 7.6 requires that the thread or lugs not override when 0.15 N·m to 0.17 N·m is applied to an R1 connector or 0.22 N·m to 0.25 N·m to an R2 connector, in each case for 5 s to 10 s. Clause 7.7 then requires that separation be possible with an unscrewing torque no greater than 0.35 N·m.

Tightening the collar to improve retention will eventually cause a failure in 7.7; reducing thread engagement to make release easier will cause a failure in 7.6. Table F.1, item 17, records 35 N as the minimum axial force for maintaining the connection in normal use and notes that this is the same value as for Luer connectors, which agrees with the lower bound of 7.4. Annex E lists ease of assembly and disassembly without tools, under fingertip control and in wet or gloved conditions, among the user needs that the design has to satisfy.

The assignment of reference connectors is specific to each clause. Clauses 7.1, 7.2, 7.3, 7.5 and 7.7 use the reference connectors of Figures C.1, C.3, C.8 and C.10, while 7.4 and 7.6 use those of Figures C.2, C.4, C.9 and C.11. The second group has a different lug form. Selecting the reference connector by family and sample side alone, without checking the clause, will give an invalid result.

Annex G records that R1 and R2 connectors were tested in a laboratory against the performance requirements and that all tests passed, covering unscrewing torque, separation force, positive-pressure fluid leakage and negative-pressure leakage. That record is part of the design confirmation summary; it does not replace type testing of a manufacturer's own product against Clause 7.

ISO 80369-2:2024, clause-by-clause series. Part 5 of 7.

Series index · ISO 80369-2 products