Subclause 6.2 applies to respiratory devices and accessories subjected to pressures between 15 kPa and 600 kPa. The R2 cone shall conform to Figure B.3 and Table B.3, and the R2 socket to Figure B.4 and Table B.4. As with R1, conformity is checked by verifying the dimensions and tolerances of the normative Annex B, and all dimensions are in millimetres unless otherwise stated.
A 6 % taper with a different envelope
R2 uses a 6 % nominal taper. The cone angle is given as an auxiliary dimension, 3.44° in Table B.3 and 3.437° in Table B.4, both printed in parentheses and therefore not acceptance criteria under Clause 3.1. The inner cone angle of the R2 cone, which is not marked on the drawing, is 0° to 1.0°. The bearing flank angles at separation differ between the two sides: 24.0° to 26.0° for the internal thread of the cone and 27.5° to 30.0° for the external thread of the socket.
The diameters are measured at defined distances from a datum. On the cone, Ød is the outside diameter at 0.500 mm from the tip (basic dimension), 8.164 mm to 8.266 mm; Øg is the large-end outside diameter at 7.000 mm from the tip, 8.554 mm to 8.656 mm. On the socket, ØD is the opening inside diameter at 0.500 mm from the open (large) end, 8.385 mm to 8.487 mm, and ØG is the small-end inside diameter at 7.000 mm from that end, 7.995 mm to 8.097 mm. Cone length e is 7.366 mm to 7.620 mm and socket cone depth E is 7.874 mm to 8.128 mm.
| Symbol | Description | R2 cone, Table B.3 | R2 socket, Table B.4 |
|---|---|---|---|
| a / A | Cone angle, 6 % nominal taper, auxiliary | (3.44°) | (3.437°) |
| b2 / B2 | Thread bearing flank angle at separation | 24.0° to 26.0° | 27.5° to 30.0° |
| Ød / ØD | Diameter at 0.500 mm from tip / from open end | 8.164 to 8.266 | 8.385 to 8.487 |
| Øg / ØG | Diameter at 7.000 mm from tip / from open end | 8.554 to 8.656 | 7.995 to 8.097 |
| e / E | Cone length / socket cone depth | 7.366 to 7.620 | 7.874 to 8.128 |
| Øf | Cone tip inside diameter | 5.300 to 5.609 | — |
| Øh / ØH | Locking thread major diameter | 11.927 to 12.181 | 11.543 to 11.797 |
| Øj / ØJ | Locking thread minor diameter | 10.657 to 10.881 | 10.301 to 10.555 |
| m / M | Thread crest width | 0.554 to 0.614 | 1.082 to 1.142 |
| n / N | Thread root width | 1.161 to 1.221 | 1.799 to 1.859 |
| p / P | Double-start right-hand thread, nominal pitch | 2.413 to 2.667 | 2.460 to 2.620 |
| Øu / ØU | Fluid cavity diameter | 3.375 to 3.629 | 3.375 to 3.629 |
| Øw / ØW | Inside diameter of the smallest cylinder over the collar | 18.340 to 18.872 | 18.340 to 18.872 |
| r / R | Radius or chamfer at tip / socket entrance | 0.000 to 0.250 | 0.000 to 0.250 |
Values transcribed from Tables B.3 and B.4 of the GB/T XXXXX.2 draft (ISO 80369-2:2024, MOD). The published standard should be consulted for the complete tables and their footnotes.
Thread form and the collar clearance condition
The cone side carries a double-start right-hand internal thread with a nominal pitch of 2.413 mm to 2.667 mm; Table B.3 notes a 5 mm lead. Dimension s2, the thread length at the bottom of the cone locking connector, is 7.493 mm to 7.747 mm, the same range as g2, the distance from the front of the gripping surface to the inner bottom face of the thread or cone. On the socket, S2 is 2.565 mm to 3.250 mm and T, from the end face to the start of the gripping structure, is 7.569 mm minimum.
The Øw / ØW condition is more demanding for R2 than for R1. The smallest cylinder that fits over the outer surface of the collar has an inside diameter of 18.340 mm to 18.872 mm, and the footnote to both tables requires the specified minimum diameter to be maintained over a length of 1.00 mm and the specified maximum diameter over the length e. A gauge that checks a single cross-section will not detect a local violation of this condition.
Table B.5 gives the engagement length L of the R2 assembly as an auxiliary dimension, with values of 4.188 mm and 4.683 mm in parentheses. The draft's note refers to Figure B.5, although Figure B.5 shows the R1 assembly and Figure B.6 the R2 assembly; this is one of several cross-references in the draft that should be checked against the published text.
Why a 6 % taper is not a Luer connector
Sharing the 6 % taper slope with the withdrawn ISO 594 series does not make R2 interconnectable with a Luer connector. The mating diameters in Tables B.3 and B.4, the locking thread major diameters of 11.927 mm to 12.181 mm and 11.543 mm to 11.797 mm, the double-start 5 mm lead, and the collar clearance of 18.340 mm to 18.872 mm all differ from the ISO 80369-7 / YY/T 0916.7 geometry. Table F.2, item 8, records the requirement that the range of mating diameters be incompatible with Luer connectors and with the other small-bore connectors of the series. Annex G describes the three-dimensional CAD analysis, using all tolerances at least material condition, nominal and maximum material condition, that was used to reach this conclusion. Conformance is established by the dimension chain in Annex B, not by the taper slope alone.
ISO 80369-2:2024, clause-by-clause series. Part 4 of 7.