Clause 6 is brief. Subclause 6.1 requires that small-bore connectors for respiratory devices and accessories subjected to pressures below 150 hPa (15 kPa) conform to the dimensions and tolerances of Figure B.1 and Table B.1 for the R1 cone, and Figure B.2 and Table B.2 for the R1 socket. Conformity is checked by verifying the dimensions and tolerances specified in Annex B. All dimensions are in millimetres unless otherwise stated. Annex B is normative.
Sealing geometry
The R1 pair uses an end-face contact cone. The sealing cone angle is 58.0° to 62.0° on both sides, with a draft angle not exceeding 1.00°. The bearing flank angle of the thread form at separation is 26.5° to 28.5° on each side. Locking is by a double-start right-hand thread with a permanently attached, rotatable internally threaded collar on the cone side; the notes to Figures B.1 and B.3 describe this arrangement.
The cone tip outside diameter is 3.600 mm to 3.700 mm and the sealing cone base diameter is 4.851 mm to 4.953 mm. The socket opening inside diameter is 5.060 mm to 5.150 mm, with a cone depth of 5.207 mm to 5.461 mm and a sealing cone base depth of 3.446 mm to 3.766 mm. Together with the 58° to 62° angle, these values produce the end-face contact on which the seal depends.
| Symbol | Description | R1 cone, Table B.1 | R1 socket, Table B.2 |
|---|---|---|---|
| a1 / A1 | Sealing cone angle | 58.0° to 62.0° | 58.0° to 62.0° |
| a2 / A2 | Draft angle, maximum | 1.00° | 1.00° |
| b2 / B2 | Thread bearing flank angle at separation | 26.5° to 28.5° | 26.5° to 28.5° |
| Ød / ØD | Cone tip outside diameter / socket opening inside diameter | 3.600 to 3.700 | 5.060 to 5.150 |
| e / E | Cone depth | 5.715 minimum | 5.207 to 5.461 |
| Øf / ØF | Fluid cavity inlet diameter, maximum | 2.850 | 2.850 |
| Øh / ØH | Locking thread major diameter | 9.267 to 9.521 (internal, root) | 8.887 to 9.141 (external, crest) |
| Øj / ØJ | Locking thread minor diameter | 7.747 to 8.001 | 7.137 to 7.391 |
| m / M | Thread crest width | 0.229 to 0.381 | 0.384 to 0.444 |
| n / N | Thread root width | 0.873 to 1.083 | 1.177 to 1.237 |
| p / P | Double-start right-hand thread, nominal pitch | 1.508 to 1.668 | 1.461 to 1.715 |
| Øw / ØW | Inside diameter of the smallest cylinder that fits over the collar | 11.811 to 12.319 | 11.811 to 12.319 |
| Øx2 | Sealing cone base diameter, cone only | 4.851 to 4.953 | — |
| Øu / ØU | Fluid cavity diameter | 2.850 maximum | (1.950), auxiliary |
Values transcribed from Tables B.1 and B.2 of the GB/T XXXXX.2 draft (ISO 80369-2:2024, MOD). The published standard should be consulted for the complete tables, including the remaining symbols and their footnotes.
Thread and collar dimensions
Four thread dimensions are specified separately on each side: crest width, root width, nominal pitch and major and minor diameters. On the cone these are 0.229 mm to 0.381 mm, 0.873 mm to 1.083 mm and 1.508 mm to 1.668 mm; on the socket, 0.384 mm to 0.444 mm, 1.177 mm to 1.237 mm and 1.461 mm to 1.715 mm. Dimension c on the cone, the projection of the cone end face relative to the gripping surface, is 0.635 mm to 0.889 mm and is measured with a movable or rotatable collar fully advanced towards the connector tip. Dimension S2, the thread length at the open end of the socket, is 2.670 mm minimum, and dimension T from the connector end face to the start of the gripping structure is 5.207 mm minimum. The radius or chamfer at the socket cone entrance, R, is 0.25 mm maximum.
The Øw / ØW dimension, 11.811 mm to 12.319 mm, is the inside diameter of the smallest cylinder that can be fitted over the outer surface of the collar structure. It constrains the envelope of the assembled connector, not the sealing interface, and is relevant when a shroud, a protective cap or an adjacent component has to clear the collar.
Dead space and the design background
Annex G records that, for R1, the space between the end of the cone bore and the socket taper was considered during design and kept below 95 % of the dead space of a conventional Luer connector. Table F.1 gives the dead space between mating surfaces as 0 mm³ to 104 mm³ and notes that it should preferably not exceed 1.25 times the value for the YY/T 0916.7 connector. Shallow dead space is also listed as a user need in Annex E. These are informative design records; the normative requirements for R1 remain the dimensions in Tables B.1 and B.2.
Measurement
Clause 6.1 says only that conformity is checked by verifying the dimensions and tolerances of Annex B. It does not prescribe measuring instruments or measurement uncertainty. In practice the cone angle, the draft angle and the thread flank angle are profile measurements, the diameters and cone depths are gauge or coordinate measurements, and the crest and root widths are separate inspection items. The auxiliary dimension ØU in Table B.2 is printed in parentheses and, under the definition in Clause 3.1, is recorded without being used for acceptance. Figure B.5 shows the mated pair, from which the relative positions of the sealing face, collar and gripping structure can be read.
ISO 80369-2:2024, clause-by-clause series. Part 3 of 7.