Part 1 of 7  ·  Clauses covered: Clauses 1, 3, 4; Figures B.5, B.6

ISO 80369-2:2024 specifies the design and dimensions of two small-bore connectors for respiratory applications. It was published in July 2024 as Edition 1, corrected in June 2025, and runs to 50 pages under ISO/TC 210 (ICS 11.040.10 and 11.040.20). The document is narrower than its title suggests. It does not specify requirements for the medical devices or accessories that carry the connectors; those are set in device-specific standards, and where none exists the general requirements of ISO 80369-1 may be used as guidance.

Clause 1. The pressure boundary between R1 and R2

Two connectors are defined. R1 is intended for devices and accessories subjected to pressures up to 15 kPa, with a breathing system given as the example. R2 is intended for pressures between 15 kPa and 600 kPa, with oxygen therapy tubing given as the example. The pressure referred to is the pressure available at the source to which the medical device is connected, not the instantaneous pressure at the connector itself.

The boundary is therefore an application boundary, not a dimension or a material class. It determines which of the two dimensional families in Annex B applies, which set of reference connectors in Annex C is used for testing, and which of the two design-input tables in Annex F is relevant. Three further notes in Clause 1 state that the intended application does not exclude the use of other connectors within that application, that requirements for alternative connectors are given in ISO 80369-1, and that the document does not specify the devices or accessories themselves.

R1R2
Application pressureup to 15 kPa (< 150 hPa)15 kPa to 600 kPa
Example in Clause 1breathing systemoxygen therapy tubing
Sealing geometryend-face contact cone, 58° to 62°6 % nominal taper
Fluid path diameterup to 2.850 mm3.375 mm to 3.629 mm
Dimensional figuresFigures B.1, B.2, B.5Figures B.3, B.4, B.6
Design-input tableTable F.1Table F.2

Design inputs should not be confused with the application boundary. Table F.1 records a single-fault pressure of 125 hPa and a negative pressure of −200 hPa for R1, while Table F.2 records 0 kPa to 10 000 kPa for R2 as the single-fault condition of a medical gas pipeline system. These are the conditions a connector design is expected to survive; the 15 kPa and 600 kPa figures in Clause 1 decide which connector is selected. Annex A explains that the test pressures of Clause 7.1 were chosen as the worst-case pressures under single-fault conditions for the two systems.

Clause 3. Terms and definitions

Clause 3 defines seven terms. Everything else is inherited from ISO 80369-1, ISO 14971, ISO 80369-7 and ISO 80369-20, and is listed in the alphabetized index of Annex I. The 2025 corrected version of the standard uses cone for a connector with an external sealing surface and socket for one with an internal sealing surface, replacing the earlier male and female terminology. The definition of cone explicitly states that the sealing surface need not be conical, so the assignment is made by the location of the sealing surface rather than by appearance.

SubclauseTermDefinition in brief
3.1auxiliary dimensionA dimension derived from other dimensions, given for reference only (ISO 10209:2022, 3.3.2).
3.2breathing systemThe gas pathway through which gas flows to or from a patient during artificial ventilation or respiratory therapy, continuously or intermittently connected to the patient's airway.
3.3coneConnector with an external sealing surface; the surface need not be conical; formerly called a male connector.
3.4medical gas pipeline systemSupply system, monitoring and alarm system, and distribution system with terminals at points where medical gas or vacuum is required.
3.5normal useOperation, including routine inspection and adjustment, and preparation according to the instructions for use, by any user. It also covers maintenance, service and transport, which distinguishes it from intended use.
3.6socketConnector with an internal sealing surface; formerly called a female connector.
3.7userPerson who operates or handles a medical device; a device may have more than one user.

The definition in 3.1 has a direct consequence for inspection. Values printed in parentheses in Annex B, such as the R1 socket fluid path diameter of 1.950 mm in Table B.2, the 6 % taper half-angles of 3.44° and 3.437° in Tables B.3 and B.4, and the engagement lengths of 4.188 mm and 4.683 mm in Table B.5, are auxiliary dimensions. They are recorded but are not acceptance criteria.

Clause 4. Non-interconnectability

The normative text of Clause 4 is a single sentence: small-bore connectors manufactured to this document shall conform to the requirements of ISO 80369-1. The five notes carry the rest of the information. They point to Annex C for the reference connectors used to evaluate non-interconnectability, to Annex D for medical devices and their attributes, to Annex E for usability requirements, to Annex F for design requirements, and to Annex G for the summary of the design assessment carried out under ISO 80369-1.

This division follows the architecture of the whole series. Part 1 defines the method for evaluating whether connectors of different applications can be physically mated, based on interference analysis at maximum and least material conditions. Parts 2, 3, 5, 6 and 7 fix the geometry for their respective applications. Because the designs in these parts have already been through the non-interconnectability evaluation, using a connector specified in the series is accepted as objective evidence, and verification can be carried out by document review. A manufacturer who adopts an alternative connector design instead takes on the full engineering analysis, risk assessment, marking and instructions-for-use obligations set out in Part 1.

The Chinese draft used for this reading is a modified adoption (MOD). Its foreword records an editorial change from "6.1" to "5.1" in the note referring to the Part 1 design assessment, which reflects the renumbering between ISO 80369-1:2018 and ISO 80369-1:2025. A compliance file should cite the Part 1 edition it was prepared against, since the same requirement carries two different clause numbers in the two editions.

The two connector families in assembly

Figures B.5 and B.6 show the mated pairs. The difference between the families is visible in the sealing cone, the collar and the grip structure, but conformance is established against the dimension tables of Annex B rather than by comparison with a drawing.

Figure B.5 from the draft standard
Figure B.5. Assembled R1 small-bore connector pair. Reproduced from GB/T XXXXX.2 (draft), Annex B.
Figure B.6 from the draft standard
Figure B.6. Assembled R2 small-bore connector pair. Reproduced from GB/T XXXXX.2 (draft), Annex B.