Annex C is normative. It specifies the reference connectors against which R1 and R2 samples are tested under Clause 7 and against which non-interconnectability is evaluated under Clause 4. Fourteen figures are given, C.1 to C.14, split into an R1 group and an R2 group. The reference connector is a measurement datum, and its material and finish are specified accordingly.
C.1. General requirements for reference connectors
Reference connectors shall be made of a corrosion-resistant rigid material with a flexural or tensile modulus of elasticity greater than 3 433 MPa (35 000 kg/cm²) and a surface roughness Ra not exceeding 0.8 μm. Unless otherwise specified, sharp edges and sharp corners have a radius not greater than 0.15 mm, a manufacturing allowance stated in the notes to the individual figures. Two symbols are used throughout: P for pitch and V for the lead of the double-start thread.
These figures can be compared with the 700 MPa threshold in Clause 5, which applies to the product connector. A reference connector is repeatedly mated and unmated across the seven performance tests, so any deformation or wear of its own surfaces would be read as a change in the sample. The high modulus and the surface finish requirement keep that error small.
C.2 and C.3. Two groups of R1 and R2 reference connectors
The figures are not fourteen variants of the same part. They divide first by connector family, then by the side of the sample being tested, and then by the group of tests for which the reference connector is intended.
| Group | R1 | R2 | Used for |
|---|---|---|---|
| First group, socket reference connectors for cone samples | Figure C.1 | Figure C.8 (locking) | 7.1, 7.2, 7.3, 7.5, 7.7 and non-interconnectability |
| Second group, socket reference connectors for cone samples | Figure C.2 | Figure C.9 | 7.4 axial load separation and 7.6 overriding |
| Cone reference connector assemblies for socket samples | Figures C.3, C.4 | Figures C.10, C.11 | Same two groups respectively |
| Conical inserts | Figure C.5 | Figure C.12 | Assembly of the cone reference connectors |
| Standard connectors used with the inserts | Figures C.6, C.7 | Figures C.13, C.14 | First and second groups respectively |
The cone reference connectors for testing socket samples are assemblies. Figure C.3 is built by joining the conical insert of Figure C.5 to the standard connector of Figure C.6, and Figure C.4 by joining the same insert to Figure C.7. The R2 equivalents are Figure C.10 (insert C.12 plus C.13) and Figure C.11 (insert C.12 plus C.14). The captions of C.3, C.4, C.10 and C.11 give the key for the two numbered components. The R2 insert of Figure C.12 is used for leakage, unscrewing torque, stress cracking, axial load separation and overriding tests, so the R2 cone reference connectors differ from the R1 arrangement in how the test groups are split.
Each clause of Clause 7 names the applicable figures. For a cone sample the choice is between C.1 and C.2, or C.8 and C.9, depending on whether the test belongs to the first or second group. For a socket sample it is between C.3 and C.4, or C.10 and C.11. The figure captions state the tests for which each connector is intended, and those captions are the practical index to the annex.
Reading the figures
The drawings are dimensioned in millimetres and carry notes on edge radii and on the P and V symbols. They reproduce the geometry of the product connectors of Annex B at the material condition needed for a datum, so a dimension that appears in both places should not be assumed identical without checking the relevant table. The draft states in its foreword that the appendix is adopted from ISO 80369-2:2024 as a normative annex, and the published standard remains the authority for the dimension values.
One textual point in the draft should be checked. The note to Figure C.9 says that the radius of sharp edges and sharp corners of all lugs is greater than 0.15 mm, whereas the corresponding notes to Figures C.1, C.2 and C.8 say not greater than 0.15 mm. Given the general requirement in C.1, the Figure C.9 wording appears to be a drafting error.
ISO 80369-2:2024, clause-by-clause series. Part 6 of 7.