Clauses 2 and 5 of ISO 80369-2:2024 are short, and most of their practical content sits in the reference list and in the informative rationale of Annex A. Together they establish which test methods are admissible and why the material threshold is a stiffness requirement rather than a strength requirement.
Clause 2. Normative references
Five standards are normatively referenced in the published document: ISO 6892-1:2019 for metallic materials tensile testing, ISO 14971:2019 for risk management, ISO 80369-1 for general requirements, ISO 80369-7:2021 for intravascular or hypodermic (Luer) connectors, and ISO 80369-20:2015 for common test methods. Four further test methods, ISO 178, the ISO 527 series, ASTM D638-22 and ASTM D790-17, are cited for determining the flexural or tensile modulus of plastics. There are no manufacturing requirements anywhere in the document; Annex H records that the standard covers the IMDRF essential principles only in part for this reason.
The Chinese draft GB/T XXXXX.2 is a modified adoption, not an identical one. Its foreword lists the substitutions made to suit Chinese technical conditions, and they matter when evidence is assembled for a domestic submission.
| ISO 80369-2:2024 | GB/T XXXXX.2 draft | Adoption | Used in |
|---|---|---|---|
| ISO 6892-1:2019 | GB/T 228.1—2021, Metallic materials — Tensile testing — Part 1: Method of test at room temperature | MOD | Clause 5, metallic materials |
| ISO 14971:2019 | GB/T 42062—2022, Medical devices — Application of risk management | IDT | Annexes D and G, risk acceptability |
| ISO 80369-1:2018 | GB/T XXXXX.1—XXXX, Part 1: General requirements (marked IDT ISO 80369-1:2025) | IDT | Clause 4, Annex G |
| ISO 80369-7:2021 | YY/T 0916.7—2024, Part 7: Intravascular or hypodermic connectors | IDT | Terminology, non-interconnectability baseline |
| ISO 80369-20:2015 | YY/T 0916.20—2019, Part 20: Common test methods | IDT | Clauses 7.1 to 7.7 |
| ISO 178 / ISO 527 / ASTM D638-22 / ASTM D790-17 | Retained, with GB/T 9341—2008 and the GB/T 1040 series given as informative notes | informative notes | Clause 5, plastics |
Other editorial changes in the draft include informative substitutions for ISO 4135:2022, ISO 7369-1:2016, IEC 60601-1:2005, IEC 62366-1:2015, ISO 18562, ISO 80369-3 and ISO 80369-5; the deletion of one note in Clause 4; the deletion of the design criterion concerning incompatibility with ISO 80369-4; and the removal of the terms nominal and responsible organization from the Annex I index, with the term socket added. A single set of measured data can serve both an ISO-based and a GB/T-based submission, but the citation lists and clause numbers have to be maintained separately.
Clause 5. Material requirements
Clause 5 requires R1 and R2 connectors to be made of material having a modulus of elasticity greater than 700 MPa in flexure or in tension. Surfaces that are not needed to ensure the non-interconnectability characteristics are exempt. Conformity is checked by testing in accordance with ASTM D638-22, the ISO 527 series, ASTM D790-17 or ISO 178 and, for metallic materials, ISO 6892-1:2019; Annex A adds that the modulus values discussed there are determined by the ISO or ASTM methods.
The 700 MPa threshold is a condition for the interference analysis, not a load rating. The non-interconnectability method in Part 1 evaluates whether any combination of maximum and least material conditions between two connector designs could permit a physical connection. That calculation assumes the parts hold their shape under assembly force, axial tension and incidental bending. A material soft enough to deform would invalidate the geometric argument on which the whole series rests. The exemption in Clause 5(b) follows from the same logic: a soft seal or a grip surface that plays no part in preventing a misconnection does not enter the calculation.
Annex C sets a much higher figure for the reference connectors used in testing: a flexural or tensile modulus greater than 3 433 MPa (35 000 kg/cm²), with surface roughness Ra not exceeding 0.8 μm and a corrosion-resistant rigid material. The reference connector is the measurement datum, not the specimen. It is assembled and disassembled repeatedly across seven performance tests and fourteen figure numbers, so its own wear and deformation have to be small compared with the effects being measured on the sample.
| Requirement | Applies to | Value | Source |
|---|---|---|---|
| Elastic modulus, flexural or tensile | R1 and R2 product connectors, surfaces needed for non-interconnectability | > 700 MPa | Clause 5 |
| Elastic modulus, flexural or tensile | Reference connectors used for testing | > 3 433 MPa | Annex C |
| Surface roughness | Reference connectors | Ra ≤ 0.8 μm | Annex C |
| Sharp edge and corner radius | Reference connectors | ≤ 0.15 mm | Annex C |
Annex A. Rationale for the material clause
Annex A is informative and its subclause numbering mirrors the normative clauses, so the sequence is not continuous. On the origin of the series it records that a CEN task group proposed in 2000 that connectors for different medical applications be made non-interconnectable by design, keeping the Luer connector for devices entering the vascular system or for hypodermic use. The connectors specified here are intended for respiratory devices, and manufacturers and responsible organizations are encouraged to report their experience to the technical committee for future revisions.
On material testing, Annex A explains why four methods are listed side by side. ASTM D638-22 and the ISO 527 series differ in fundamental respects, including test speed and the way modulus is calculated. An interlaboratory comparison covered four unfilled, unreinforced and unpigmented thermoplastics, namely polycarbonate, polybutylene terephthalate, acrylonitrile butadiene styrene and high-impact polystyrene. Yield tensile stress, tensile stress at break, elongation at yield and elastic modulus were analysed after removal of outliers. Agreement between the ISO and ASTM data varied by material and property, but the elastic modulus values were closely similar. That result is the basis for allowing either family of methods in Clause 5.
The same annex states that the test pressures selected for the leakage clause represent the worst-case pressure that could arise under a single-fault condition for an R1 connector in a breathing system and an R2 connector in a medical gas pipeline system. This connects Clause 2, Clause 5 and Clause 7.1 more directly than the clause text alone suggests.
Two textual points in the draft are worth checking against the published standard. The foreword refers to ISO 80369-1:2018 while the normative reference list marks GB/T XXXXX.1 as identical to ISO 80369-1:2025. In Annex A, the subclause headed "pressure decay leakage" is numbered 6.1, which follows an earlier draft structure in which performance requirements formed Clause 6; in the published standard and in the draft's own Clause 7 it is 7.1.
ISO 80369-2:2024, clause-by-clause series. Part 2 of 7.