chinatic.org / ISO 80369 Series
ISO 80369-2:2024 · Clause-by-clause reading

ISO 80369-2:2024: Rationale and Informative Annexes (A, D to I)

Part 7 of 7  ·  Clauses covered: Annexes A, D, E, F, G, H, I

Of the nine annexes in ISO 80369-2:2024, only B and C are normative. The remaining seven explain where the requirements came from, which devices use the connectors, what the design was expected to achieve, and how the design was assessed. They are the main source for the reasoning behind the dimension tables and the performance limits.

Annex A. Rationale and guidance

Annex A is written for readers familiar with the subject but not involved in drafting the standard, and its subclause numbers mirror the normative clauses, so the numbering is not continuous. It records that a CEN task group proposed in 2000 that connectors for different medical applications be non-interconnectable by design, retaining the Luer connector for intravascular and hypodermic use. Manufacturers and responsible organizations are invited to report experience to the technical committee for future revisions. Two passages are of direct technical use: the single-fault rationale for the Clause 7.1 test pressures, and the interlaboratory comparison of ASTM and ISO modulus measurements on PC, PBT, ABS and HIPS that supports the choice of test methods in Clause 5.

Annex D. Medical devices and their attributes

Table D.1 lists devices and accessories with connections in this application, grouped under seven headings: breathing system connections, sidestream gas monitor connections, oxygen source connections, oxygen conserving system connections, humidifier system connections, negative pressure connections, and connections for delivering liquids or drugs. Each entry is assessed for maximum pressure class (A below 150 hPa or B above 150 hPa), air or oxygen, barotrauma and hypoxia risk control, infusion fluid type, pressure range, flow range, patency dynamic indicator, whether locking is required, and whether injection is required. Pressure measuring lines are listed below 125 hPa at flows below 0.05 L/min; sidestream gas sampling lines run from −400 hPa to 125 hPa at 0.02 L/min to 0.5 L/min; oxygen delivery source outlets are listed below 6 000 hPa at up to 30 L/min. Three footnotes define the term port, state that some connections are designed for specific drug nebulizers, and describe ports with a rigid connector and a resilient socket.

Annex E. Usability requirements

The use environment covers hospitals, clinic settings, home care, medical transport and emergency care, in each case at a location with a medical gas outlet. The stated ambient temperature range is −40 °C to +60 °C, justified by field use in emergency medical services and outdoor use by home care patients. Three attributes of the use scenario are recorded: usability under stress where a user forces an assembly without reading a label, several fluids in close proximity with gloved hands, and the proximity of other devices carrying connectors.

Seven user needs are listed: minimal training for clinical users; operation without tools and easy assembly and disassembly under fingertip control, in particular when wet or gloved; no misconnection with other small-bore connectors in the same environment; no leakage in normal use; security of the connection with no inadvertent disconnection; shallow dead space; and compatibility with disinfection, decontamination, sterilization and reprocessing. The last three of these are reflected in the mechanical clauses of Clause 7, the dead space figures in Annexes F and G, and the material clause respectively.

Annex F. Summary of design requirements

Tables F.1 and F.2 set out 33 design inputs for R1 and R2. They are informative, and several values differ from the application and test pressures of Clauses 1 and 7 because they describe the design envelope rather than a conformity test.

ItemR1, Table F.1R2, Table F.2
Fluid typeliquids and gasesgas
Pressure range0 hPa to 150 hPa; 125 hPa single fault; −200 hPa negative0 kPa to 10 000 kPa, medical gas pipeline single fault
Flow range0 L/min to 15 L/min0 L/min to 100 L/min
Cone through-bore1.00 mm to 2.95 mm5.11 mm to 5.36 mm
Socket through-bore1.00 mm to 2.70 mm4.57 mm to 4.83 mm
Temperature range−40 °C to +60 °C−40 °C to +60 °C
Overall designconical, tip sealconical
Minimum axial force in normal use35 N35 N
Structural material modulus> 700 MPa> 700 MPa
Dead space between mating surfaces0 mm³ to 104 mm³—

Both tables require resistance to stress cracking, absence of sharp edges, an elastic modulus above 700 MPa, no soft sealing material, no colour coding and no labelling or symbols to indicate intended use, and they leave MRI compatibility open as option b or d depending on the application. Biological compatibility is left as option a or b with a reference to the YY/T 1778 series for respiratory gas pathways in the draft. Items 29 to 32 of each table record how incompatibility with the other connector families of the series is achieved, by dimension in each case.

Annex G. Summary of the design assessment

Annex G is divided into general, design engineering analysis, design verification, design confirmation and design review. The general clause states that no patents are known for the designs and that they prevent a fluid seal or interconnection with other standard connectors in ISO 80369-1, subject to what follows in G.2. The engineering analysis section describes the two geometries: an end-face contact cone for R1 and a 6 % cone for R2, with threads added on each side of the bore so that the parts cannot easily separate and can resist axial pull. It also records that the R1 dead space was kept below 95 % of that of a conventional Luer connector.

The analysis used calculation and three-dimensional solid CAD models built with all tolerances at least material condition, nominal and maximum material condition, together with the material conditions of every connector in the series. The process itself is described in ISO 80369-1. Table G.1 reports three residual misconnections that were judged acceptable.

The first is an R1 cone with an E1 enteral socket. The minimum, average and maximum overlap of the contactable surfaces are −0.210 mm, −0.480 mm and −0.750 mm. Interference analysis under Part 1, Table B.2, shows that a full-diameter to full-diameter connection cannot be made at an interference of −0.284 mm, but that a misconnection can occur when the overlap is only −0.129 mm. Both connectors would have to be at or near their worst-case dimensions, which is considered unlikely, and even then the connection would not be maintained without applied pressure, so no fluid could pass unless the parts were held together. The risk is therefore recorded as acceptable.

The second and third are known misconnections: an R1 socket with an S1 limb cuff cone, which would connect a breathing system to a limb cuff, and an R2 cone with an S1 limb cuff cone, which would connect a cannula or mask to a limb cuff. Risk assessment in each case judged the connection acceptable. These are conclusions under the risk management framework of ISO 14971 for the connector designs themselves. They do not permit a manufacturer to design a device that mates deliberately, and the residual risk has to be reassessed for a specific device and its use scenario.

Table G.1, first case from the draft standard
Table G.1, first case. Potential misconnection between an R1 cone and an E1 enteral socket. Reproduced from GB/T XXXXX.2 (draft), Annex G.
Table G.1, second case from the draft standard
Table G.1, second case. Known misconnection between an R1 socket and an S1 limb cuff cone. Reproduced from GB/T XXXXX.2 (draft), Annex G.
Table G.1, third case from the draft standard
Table G.1, third case. Known misconnection between an R2 cone and an S1 limb cuff cone. Reproduced from GB/T XXXXX.2 (draft), Annex G.

Annex H and Annex I

Table H.1 maps four IMDRF essential principles to Clauses 4, 5 and 6: 5.3.3, 5.5.1, 5.5.2 b) and 5.5.5. Each row says that the principle is covered only in part, because the document gives no manufacturing requirements; what it does address is the effectiveness of the intended connection and the mechanical prevention of misconnections that would present an unacceptable risk. A principle not listed in Table H.1 is not addressed by the document. Annex I is an alphabetized index of defined terms with the source of each definition; the draft adds socket and removes nominal and responsible organization.

Editorial notes on the draft

The Chinese draft used for these readings is a consultation text dated October 2026, and a few internal cross-references and version references do not agree with one another. Besides the Clause 4 renumbering of Part 1 and the Annex A subclause number discussed in Part 2 of this series, the draft's Annex H cites IMDRF/GRRP WG/N47:2018 while bibliography entry 10 gives N47:2024; the design confirmation summary in Annex G refers to "the performance requirements in Clause 5" where the performance requirements are in Clause 7; items 29a and 29b of Tables F.1 and F.2 ask how incompatibility with GB/T XXXXX.2 is achieved, which is the draft's own part number; the note to Table B.5 refers to Figure B.5 rather than B.6; and the note to Figure C.9 reverses the sharp-edge radius condition. These are textual observations on a consultation draft. The published ISO 80369-2:2024, including its June 2025 corrected version, and the final GB/T text are the authorities for any conformity decision.

ISO 80369-2:2024, clause-by-clause series. Part 7 of 7.

Series index · ISO 80369-2 products