IEC 80369-5:2016: Rationale and Informative Annexes (A, E, F, G, H)

Part 7 of 7  ·  Clauses covered: Annexes A, E, F, G, H; Figure H.1; Bibliography

Seven annexes in the draft are informative: A, E, F, G, H, I and J. Three are normative: B, C and D. The informative annexes explain where the requirements came from, who uses the connectors, what the design was expected to achieve, how the design was assessed and which older connectors were withdrawn. Annexes I and J are read separately in Part 6 of this series.

Annex A. Explanation and guidance

Annex A is addressed to readers familiar with the document but not involved in drafting it, and its subclause numbers mirror the normative clauses, so the numbering is not continuous. It records that a CEN task group proposed in 2000 that connectors for different medical applications be made non-interconnectable by design, retaining the Luer connector for devices entering the vascular system or for hypodermic use. Manufacturers and responsible organisations are invited to report their experience to SAC/TC 221 for consideration when the series is revised.

Four notes carry technical content. On Clause 5.1, the S1 connector is described as widely available from many sources, already accepted in the market and non-interconnectable with the other connectors of the series, with the reservation that the unspecified locking and sealing mechanism leaves dimensions undefined. On Clause 6.2, the home healthcare exception is explained by the patient being the user. On Annex C.2, the standard connector is identified as the original manufacturer's S1 design. On I.2 and J.2, the moisture absorption of hygroscopic materials is given as the reason for preconditioning.

Annex E. Usability requirements

Non-invasive blood pressure is measured across almost every patient group, so the user population is broad. Table E.1 separates patients and primary care providers, who may be untrained and may have an impairment such as reduced vision, limited ability to handle small connectors or difficulty reading instructions, from occupational clinical users with various levels of training. Patients may have direct contact, clinical users direct or indirect contact. Users listed include the patient, physicians and physician assistants, nurses at various levels, home care providers, visiting nurses, relatives and emergency medical technicians. Neonatal patients are outside the user profile because their connectors are not yet in the document.

Many connections are expected to remain untouched during use, but the draft records the cases where disconnection occurs: personal interventions such as changing clothes or leaving a continuously monitored area, patient transport, and emergency deflation of a home-use cuff. Use environments include commercial premises such as pharmacies, homes, independent clinics, emergency and field medical settings, and hospitals, for both single measurements and continuous monitoring. The listed locations extend to operating rooms, wards, delivery rooms, intensive care units, physicians' offices, pain clinics, field hospitals, transport systems, infusion clinics and assisted care facilities.

Two ambient conditions are given: −20 °C to +50 °C, justified by field use in emergency medical services and outdoor use by home care patients, and body temperature up to 42 °C. Seven general user needs follow: ease of operation without the force needed for a Luer connection; no connection with other connectors in the use environment; no leakage in normal use so that blood pressure measurement stays accurate; connection reliability recognisable mechanically or cognitively by sight, touch or hearing; the ability to disconnect for pressure release or on device failure; materials suitable for skin contact; and decontamination, since sphygmomanometers and cuffs are typically reusable and require cleaning and sometimes disinfection or sterilisation.

Annex F. Design requirements

Tables F.1 and F.2 set out 34 criteria each, for adult or paediatric connectors and for neonatal connectors. Several entries explain values that appear in the normative clauses: the operating pressure is set at least 10 % above single-fault pressure, the leak criterion is limited to 10 % of the leakage allowed for the sphygmomanometer and cuff together, and the minimum flow is chosen so that the pneumatic system can be depressurised within 10 s. The fluid is gas, the bore is 3 mm to 4 mm for tubing and 1 mm to 5.2 mm for the connector, and the overall construction is a parallel O-ring seal or another parallel seal rather than a cone, specifically to avoid Luer compatibility. Both tables require the absence of sharp edges, biological compatibility, no colour coding and no labelling or symbols indicating intended use. Incompatibility with the Luer connectors of GB/T 1962.1 and GB/T 1962.2 and with Parts 2, 3, 6 and 7 of the series is achieved by dimension in every case.

Annex G. Design assessment

The S1 connector in Annex B has more than 20 years of clinical use for adult and paediatric blood pressure measurement, joining the bladder to the sphygmomanometer or its hose. Users found it clinically acceptable, and it is non-interconnectable with the 6 % female Luer connector used for intravascular and hypodermic applications. An assessment summary for neonatal cuff connectors and for tourniquet connectors is to be added to the annex when those connectors are introduced by amendment or revision.

The engineering analysis used three-dimensional solid CAD models built with all tolerances and material conditions for every connector described by this document and by the rest of the series. It concluded that the S1 connector is non-interconnectable with the other connectors of the series, with one residual case recorded in Table G.1: a male S1 against a male N1, N1 being the connector from YY/T 0916.6-2022. Physical testing to Annex B of Part 1 confirmed that the two do not connect. A technical report describing the CAD analysis in more detail was planned.

Design verification, confirmation and review are all treated as unnecessary under the corresponding procedures of GB/T XXXX.1-202X, because of the long and successful clinical history. Two configurations of the S1 connector were nevertheless tested, one in nylon with a nominal tensile modulus of 1200 MPa and one in metal, both widely available materials meeting 4.2. Performance tests to Clause 6 using the methods of GB/T XXXX.20 showed that the S1 design meets the requirements. The committee reviewed the assessment and concluded that there is significant objective engineering, technical and clinical evidence supporting the intended application.

Annex H. Withdrawn connectors

The connectors described in Annex H were used for decades on non-automatic sphygmomanometers. They are known to fail the non-interconnectability tests of Annex B of Part 1, so the draft excludes them from future use. A large installed base of sphygmomanometers remains, with service lives measured in decades, and those instruments will need a permanent conversion connector to adapt the withdrawn fitting to the S1 connector of Annex B. Figure H.1 and Tables H.1 and H.2 preserve the historical dimensions for that purpose.

Withdrawn male connector, Table H.1Withdrawn female connector, Table H.2
Thread5/16-24 UNF-2A internal5/16-24 UNF-2A external
Length of male plug / female projection5.25 min6.700 min
Outside diameter of male plug Øh3.495 to 5.615—
Sealing surface inside diameter ØH—5.610 to 5.730
Inside diameter of shroud Øj7.940 to 9.000—
Engagement length (L), reference(2.250)(2.250)
Fluid cavity inside diameter Øu / ØU, recommended1.651 to 2.9512.200 to 2.960
Projected width of body Øw / ØW9.500 to 11.5607.910 to 11.610

Table H.2 of the draft prints several values with commas as decimal separators, for example 3,250 and 5,710. They are read here as 3.250 and 5.710. The comma usage is a transcription artefact of the consultation text, not a different dimensioning convention.

The 5/16-24 UNF thread is the distinguishing feature of the withdrawn system. A conversion connector has to bridge that threaded interface and the bayonet-style S1 geometry of Annex B while still satisfying the non-interconnectability requirement of Clause 4.1, which is the reason the historical dimensions are retained rather than simply deleted.

Figure H.1 from the draft standard
Figure H.1. Withdrawn blood-pressure and inflation small-bore connector. Tables H.1 and H.2 give the historical dimensions. Reproduced from the GB/T XXXXX.5 consultation draft, Annex H.

Bibliography and textual observations

Seventeen references are listed. They include the Luer connector standards GB/T 1962.1 and GB/T 1962.2, GB 9706.1-2020, the other parts of the small-bore connector series, YY 9706.111 and YY 9706.112 for home healthcare and emergency medical service environments, YY 9706.230 for automated non-invasive sphygmomanometers, ISO 81060-1:2007, ISO 3040:2009, ISO 80369-2:2018, ISO 80369-20:2015, IEC 62366-1:2015, ASME/ANSI B1.1:2003 for unified inch screw threads and the CEN CR 13825 report on Luer connectors.

A consultation draft is not a conformity document, and several internal references do not agree with one another. The foreword substitutes GB/T 1040.1-2025 for ASTM D638-14 and ASTM D790-10 for ASTM D790-25, while Clause 2 still lists ASTM D638-14 and ASTM D790-25 and Clause 4.2 calls up ASTM D790-25. Two notes in Annex B say the male sealing surface Øh5 mates with ØS5, whereas Table B.2 gives ØS5 as 4.380 mm and the female bore ØH5 as 5.060 mm to 5.150 mm. Clause 6.1 uses seconds while Annex I uses minutes. Table F.2 defines design inputs for a neonatal connector that Clause 1 defers to a future amendment. Annex K of the source standard was deleted. The published IEC 80369-5:2016 with Cor 1:2017 and Cor 2:2021, and the final GB/T text, are the authorities for any conformity decision.